Input Types | K, J, E, T, B, R, N, S, C nine thermocouple types; With Dual-channel input | |
---|---|---|
Measuring range | K | - 200.0 ~ + 1370.0 ℃ (-328.0 ~ +2498.0 ℉) |
J | - 200.0 ~ + 1200.0 ℃ (-328.0 ~ +2192.0 ℉) | |
E | - 210.0 ~ + 1000.0 ℃ (-346.0 ~ +1832.0 ℉) | |
T | - 220.0 ~ + 400.0 ℃ (-364.0 ~ +752.0 ℉) | |
B | + 320.0 ~ + 1800.0 ℃ (+608.0 ~ +3272.0 ℉) | |
R | - 20.0 ~ + 1700.0 ℃ (-4.0 ~ +3092.0 ℉) | |
N | - 200.0 ~ + 1300.0 ℃ (-328.0 ~ +2372.0 ℉) | |
S | - 20.0 ~ + 1750.0 ℃ (-4.0 ~ +3182.0 ℉) | |
C | 0.0 ~ + 2300.0 ℃ (+32.0~ +4172.0 ℉) | |
Accuracy | ± 0.1% of rdg + 0.1℃ (In 23±5℃ operating environment) ± 0.1% of rdg + 0.2℃ (Out off 23±5℃ operating environment) ※Built-in cold junction compensation function. |
|
Resolution | K,J,E,T,N,C Type: 0.1℃ / 0.1℉; B,R,S Type : 0.5℃ / 0.5℉ |
|
Reading Rate | Approx 0.4 sec | |
Logging Sampling Rate | 2 second to 120 minutes (User Selectable) | |
Memory | 16,000 readings x 2CH | |
Baud Rate | Switchable 9 kinds of thermocouple type input, Hi/Lo Alarm, T1-T2, Data hold, Max/Min/Avg Functions, USB/RS-232 interface, Perpetual calendar, Data-Logging, Switchable ℃/℉, AC/DC power, Battery sign and low battery warning, Auto / Manual shutdown, Calibration function, Large LED back-light, IP66 water and dust proof. |
|
Output | a. Software with USB Interface cable b. RS-232 Output | |
Power Source | One 9 V battery or AC Adaptor(option) | |
Input Connections | Standard mini thermocouple socket x 2 | |
Operating Environment | -20~+60℃; 0~100%RH | |
Dimensions / Weight | 150 x 75 x 28 mm, Approx. 320g (battery included) |
※ About accuracy and sampling rate excludes errors generated by temperature probe.
No. | Model No. | Descriptions |
---|---|---|
① | TX-600N | Digital thermometer, 2ch. Data-Logger |
② | Temp. Probes | LP series ※Please refer to catalog P30~P34 |
③ | TU-RS232-W | RS-232 Interface cable |
④ | TU-USB-W | USB computer interface cable |
⑤ | TU-655C | Plastic carrying case |
⑥ | TU-6019 / 6029 | AC Adaptor |
No. | Model No. | Descriptions |
---|---|---|
① | TX-600N | Thermometer / Data Logge |
② | TU-USB-W | USB computer interface cable |
③ | TU-RS232-W | RS-232 Interface cable |
④ | Software CD | Software description |
⑤ | TU-6019/6029 | AC Adaptor |
⑥ | TU-609 | Battery 9V006P |
⑦ | User's manual | Multi-national user's manua |
⑧ | TU-655C | Plastic carrying case |