Model No. | TH-380 | |
---|---|---|
Measurement Range | Relative Humidity : | 0 ~ 100% RH |
Temperature : | -40 ~ +85℃,-40 ~ +185 ℉ | |
Sensors | Humidity : | Thin Polymer capacitor |
Temperature : | 100 ohm platinum RTD | |
Resolution | 0.1 % RH, 0.1℃ / 0.1℉ | |
Accuracy (at 25℃) | Humidity : | ± 2% (0 ~ 95% RH) |
Temperature : | ± 0.3℃ / ± 0.5℉ | |
Response (90%) | 30 sec. (with membrane filter) 10 sec. (without filter) |
|
Display | LCD display (4-digital), Temperature and Humidity display simultaneous. | |
Main Functions | Records MAX/MIN values, Data-Hold, ℃/℉ switchable, Auto/Manual power off, Low power warning,Offset calibration function, sensor fault warning. | |
Power Supply | One 9V battery (alkaline), or AC adaptor (optional) | |
Battery Life | Approx. 160 hours typical | |
Operating/Storage Conditions | Instrument : | 0 ~ 50℃ (+32 ~ 125℃) |
Probe : | -20 ~ 85 ℃(5 ~ +185℉) | |
Dimensions / Weight | Nstrument : | 168 × 73 × 33 mm (6.6”× 3” × 1.3”) / Approx.170g included battery |
Probe : | 20 × 220Lmm ( 0.8” × 8.7” Length) / Approx.100g | |
Cable length : | Approx. 920 mm (36”) | |
Approvals | CE, RoHS | |
Accessories | Temperature / Humidity probe (No.TH-30) | |
Latching carrying case (No.3812C) | ||
9V Battery | ||
Operation manual |
※ Specifications are subject to change without notice
Model No. | Descriptions |
---|---|
TH-30 | Replacement probe for TH-380 |
TH-3811 | AC adaptor |
TH-3812 | Plastic carrying case |
TH-3813-33 | 33%rh Standard solution |
TH-3813-75 | 75%rh Standard solution |